The Brittleness Point Temperature Tester - QT-BPT Series Brittleness Tester covers the determination of the temperature at which rubber, plastics and elastomers exhibit brittle failure under specified impact conditions. Samples to be housed in a thermostatic bath based on the reference standards ASTM D746, ISO 812, as well as and other international standards.
The cooling medium is dry ice and Nitrogen. The working temperature is from –70°C to +20°C. With this test method the test specimens are mounted onto cantilever grips and then immersed into the low temperature bath for 3 minutes (rubber), or 2.5 minutes (other materials).
The specimen is then struck with the tip of the hammers dart at a linear speed of 2 ± 0.2m/s at a lowest temperature of no occurrence of brittle failure. The sample is then checked for cracks, crevices, small holes and fragments.
|
QT-BPT100 |
QT-BPT100-1 |
Temp |
Room temp.~-70℃ |
Ambient temp. to -120℃(LN2 based) |
Temp. control |
Temperature controller, SSR and electric heater |
Temp. controller+SSR |
Temp. Resolution |
±0.01℃ |
±0.5°C |
Cooling method |
dry ice (prepared by customer) |
Liquid Nitrogen |
Striking linear speed |
2.0 ± 0.2m/s |
2.0±0.2 m/s, radius (1.6±0.1 mm) |
Pilot chute size (W×D×H) |
570 × 180 × 160 mm |
- |
Dimension(W×D×H) |
670 × 670 × 550 mm |
650 × 680 × 550 mm |
Weight |
136 kg |
~150kg |
Power |
220V 50/60Hz, 3A |
220V 50/60Hz, 3A |